PO2 MONITOR
A VERY REAL WARNING!!!
The PO2 monitor below was developed by me for my needs. I can not and do not make any claims as to its applicability or capability to monitor PO2 values. Improper monitoring of your PO2 whilst diving a rebreather can lead to your death. There may be errors in the drawing below. Get the proper data sheets and verify your own work.
When working, the Draeger oxygauges performed well. I had heard many others complain of problems, but until recently, never had one go bad. However, I found a major flaw in their design after loosing two of them on two successive dives. That flaw turns out to be a major one.
Following these failures and discovering the root cause, made me decide to make my own gauges. I knew the basic requirements as I already outlined in a previous article I wrote at www.airheadsscuba.com/sensors.html; I would require two gauges. Each gauge had to have its own supply and they had to be reliable. Last, but not least, I should be able to monitor both units simultaneously. I don't use a primary and backup unit for PO2 monitoring. I use two primaries. All the alternatives I came across cost way too much.
My Unit
|
Dual gauge housing |
Housing and sensor holders |
Access to an end-mill and drill press is all you need. I started with a 50 mm X 76 mm X 100 mm (2 inch thick X 3 inch wide X 4 inch long) aluminum bar stock. I milled out the cavities to install the batteries and circuit card. I used standard off the shelf hoses, fittings and P connector housings from www.oxycheq.com. The total cost for everything was under US$200.00. That works out to a third the price of one oxygauge or a fifth the cost of a VR3. As you can see from the photos, the housing I have settled on is an ambient pressure two section housing. Each gauge sits in its own sealed pocket and has its own battery supply. I am using normally closed magnetic reed switches to switch power. Using the normally closed contacts enable me to remove a strap with the magnets attached to activate them. I installed the old tried and true barbed fittings for the polyurethane hose. I have used them on my oxygen injection system and they have never given a moments problem. There are pros and cons in an ambient pressure housing. On the plus side it is easier to make a housing that only has to withstand 70 - 140 mb (1 or 2 psi). On the negative side a flood of the case will allow water to get into the loop and a flooded loop will of course flood the PO2 gauge housing. The LCD used is nice because a flooded housing doesn't necessarily mean a fried meter. Its circuitry is epoxy encapsulated!
|
|
Magnetic in place |
Switch and magnet for backlight |
|
|
Backlight |
Pressure pot made from pressure cooker |
|
| Circuit Descriptions
Basic PO2 gauge schematic The basic PO2 monitor is drawn right from www.ppo2.com . It is a meter, a sensor, and a potentiometer. It is an oxygen analyzer. If you do not need an audible or visual alarm, you are in business. I used a pressure pot and ran the PO2 from 0,21 ata to 1,95 ata and the oxygauge and my unit read the same all the way up! Without an alarm circuit, the power drain from the LCD meter is only about 240uA. With an alkaline 9 volt supply, a power switch really isn't necessary. A three months run time should be normal. |
||
![]()
|
Adding an alarm circuit however, increases significantly the
power requirements for the circuitry. It also adds the requirement to add an on/off
switch. I found a real nice device made by Texas Instruments (TI) which has a dual op amp
and dual comparator stage in a 14 pin DIP. Real convenient. One of the amplifier sections is set up to give a gain of 10 to have some operating room for the comparator. The two comparator sections are set up as a window comparator. The resistor divider network comprised of R03, R04, and R05 establish the thresholds for PO2 low and PO2 high. The values selected give an alarm at 0,345 PO2a and 1,55 PO2ata. D01 is a 2,5 VDC micro power shunt reference diode. Using this diode should eliminate variations in battery voltage on the alarm settings.
|
|
![]() |
If you do not need a beeping alarm and are willing to settle
for a constant audible buzzer, the only problem you will find is the space to put the
acoustic beeper and IC's. A constant alarm is also not as attention getting as an
interrupted beep. The signals are taken to + supply because the comparator sections
are open collectors. Another option I have been giving some consideration is to replace the two options to the left and put a solenoid in their place. This would control diluent though a 15 lpm orifice. In case of a low or high PO2, this would add diluent to the loop. The danger here is when you are running a low FO2 in this supply you may become hypoxic. i.e. FO2 of 10% shallow will insure your demise. The alarm indication in this case is bubbles escaping from the loop. I will also add a switch to disable this feature for deco. It will be a deliberate two step process, i.e. Pull the strap with the magnet in it and place it over the switch. |
|
If you want a pulsed output, you will need a timer. The old standby NE555 timer is not a good selection. Its power requirements in standby are almost 8 ma. This is way to high. Again, TI makes a low power equivalent in the TLC555. Its standby power is less than 1 ma. This is still high, but acceptable. Of course, when the acoustic alarm goes off the peak power is in the order of 10's of milliamp. |
||
![]() |
The whole schematic and all the options are seen by clicking on the picture below. |